Research Journal of Biological Sciences

Year: 2008
Volume: 3
Issue: 12
Page No. 1381 - 1386

AV Interval and Cardiac Output in Patients with Implanted DDD Pacemaker

Authors : Fariborz Akbarzadeh , Mehrnoush Toufan and Nasrin Afsarpour

Abstract: Our aim was to compare the Cardiac Output (CO), Filling Time (FT) and Myocardial Performance Index (MPI) derived optimal atrioventricular delay (AVD) and to analyze systolic and diastolic performance at every optimal AVD. We studied 32 patients with implanted DDD pacemaker, from implantation time to 6 months following PM implantation, in Tabriz Shahid Madani Heart Center. The evaluation was performed during AV sequential pacing with different programmed AVDS ranged from 100-200 ms by steps of 20-30 ms. At every AVD, the following parameters were measured: FT, mitral VTI, ET, aortic VTI, ICT and IRT. CO and FT derived optimal AVDs were significantly different (146�37 and 126�35 ms, respectively), but their difference with MPI derived optimal AVDs was not significant (130�28 ms). ICT/ET was similar at CO, FT and MPI derived optimal AVD (0.24�0.10, 0.22�0.05 and 0.20�0.07, respectively). IRT/ET ratio was similar at CO, FT and MPI derived optimal AVDs (0.46�0.14, 0.45�0.10 and 0.42�0.10, respectively).

How to cite this article:

Fariborz Akbarzadeh , Mehrnoush Toufan and Nasrin Afsarpour , 2008. AV Interval and Cardiac Output in Patients with Implanted DDD Pacemaker. Research Journal of Biological Sciences, 3: 1381-1386.

Design and power by Medwell Web Development Team. © Medwell Publishing 2024 All Rights Reserved